Home / Recalls / Medical Devices
Maquet Cardiovascular Us Sales, Llc recalls Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86682003 Product Usage: The Getinge 86 Series Medica
Recalled June 19, 2019 · FDA recall Z-1778-2019 · Maquet Cardiovascular Us Sales
Hazard
Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process. · FDA Class II
Units
40 units
Sold at
Worldwide Distribution - US Nationwide GA, IL, IN, MI, NC, OR, PA, VA Foreign: Australia, Canada, China, France, Germany, India, Japan, Korea, Netherlands, Saudi Arabia, South Africa, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, Ukraine
Description
Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process.
Product
Getinge 86-Series-Medical Washer-Disinfector Part Number: S-86682003 Product Usage: The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.
Manufacturer(s)
Maquet Cardiovascular Us Sales, Llc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →