RecallWatchUSA

Home / Recalls / Medical Devices

Maquet Cardiovascular Us Sales, Llc recalls Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH

Recalled July 29, 2015 · FDA recall Z-2097-2015 · Maquet Cardiovascular Us Sales

Hazard
There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision. · FDA Class II
Units
12 units
Sold at
Worldwide Distribution - US Nationwide Distribution and to the countries of : France, Australia, Belgium, Brazil, Switzerland, Chile, China, Germany, Spain, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Japan, Kuwait, Kazakhstan, Mexico, Netherlands, Norway, Russia, Sweden, Turkey, Taiwan and South Africa.

Description

There is a potential issue that could result to collision between the C-arm of the angiography system and the MAGNUS OR table system. After switching on the angiography-system, in rare cases, MAGNUS column provides wrong position data for the longitudinal movement leading to a collision.

Product
Magnus Hybrid OR Table Column (1180.01B2) used in conjunction with Siemens Artis Zeego Angiography System. Maquet GmbH & Co.
Manufacturer(s)
Maquet Cardiovascular Us Sales, Llc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov