RecallWatchUSA

Home / Recalls / Medical Devices

MAQUET GMBH recalls Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F

Recalled September 3, 2025 · FDA recall Z-2420-2025 · MAQUET

Hazard
Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel. · FDA Class II
Units
315 units (8 US, 307 OUS)
Sold at
Worldwide - US Nationwide distribution in the states of KY, NY, TX and the countries of Algeria, Australia, Austria, Bulgaria, Canada, China, Colombia, Denmark, Egypt, Estonia, Finland, France, Germany, Ireland, Israel, Italy, Japan, Jordan, Korea, Republic of, Kuwait, Lebanon, Morocco, Netherlands, Norway, Poland, Portugal, Romania, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom.

Description

Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.

Product
Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions.
Manufacturer(s)
MAQUET GMBH
Made in
Germany

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

MAQUET GMBH recalls Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F | RecallWatchUSA