RecallWatchUSA

Home / Recalls / Medical Devices

Mar Cor Purification recalls Mar Cor Purification CWP Reverse Osmosis System. Models 101, 102, 104, and 106.

Recalled March 19, 2014 · FDA recall Z-1201-2014 · Mar Cor Purification

Hazard
There exists a potential situation with the crimp connection at the pump head assembly. The potential exists that the reverse osmosis pump head wiring could build resistance resulting in resistive heating of the surrounding components. If this occurs, the pump head assembly may experience melting of the plastic components and potentially create smoke/fire and shut down of the unit. · FDA Class II
Units
130
Sold at
nationwide. No distribution outside the US.

Description

There exists a potential situation with the crimp connection at the pump head assembly. The potential exists that the reverse osmosis pump head wiring could build resistance resulting in resistive heating of the surrounding components. If this occurs, the pump head assembly may experience melting of the plastic components and potentially create smoke/fire and shut down of the unit.

Product
Mar Cor Purification CWP Reverse Osmosis System. Models 101, 102, 104, and 106.
Manufacturer(s)
Mar Cor Purification

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov