Home / Recalls / Medical Devices
Mar Cor Purification recalls Mar Cor Purification Millenium HX Portable Reverse Osmosis System. Model 3025975. Product Usage: Water Purific
Recalled March 26, 2014 · FDA recall Z-1213-2014 · Mar Cor Purification
Hazard
High inlet water pressure beyond specification and the solenoid valve does not meet application which could result in a delay in treatment. · FDA Class II
Units
232 units
Sold at
USA Nationwide Distribution including Puerto Rico.
Description
High inlet water pressure beyond specification and the solenoid valve does not meet application which could result in a delay in treatment.
Product
Mar Cor Purification Millenium HX Portable Reverse Osmosis System. Model 3025975. Product Usage: Water Purification System is intended to be used as a dialysis accessory to produce water through reverse osmosis for use with hemodialysis equipment..
Manufacturer(s)
Mar Cor Purification
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →