RecallWatchUSA

Home / Recalls / Medical Devices

Mc3 Inc recalls NautilusTM ECMO Oxygenator with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, re

Recalled March 27, 2024 · FDA recall Z-1309-2024 · Mc3

Hazard
May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer · FDA Class II
Units
299 units
Sold at
US Nationwide distribution.

Description

May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer

Product
NautilusTM ECMO Oxygenator with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Product Number: (1) 48145; (2) 48145E
Manufacturer(s)
Mc3 Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Mc3 Inc recalls NautilusTM ECMO Oxygenator with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, re | RecallWatchUSA