RecallWatchUSA

Home / Recalls / Medical Devices

Mckesson Medical-Surgical Inc. Corporate Office recalls Abbott Point of Care Control Control i-STAT¿ Level 3 1.7 mL

Recalled March 27, 2024 · FDA recall Z-1315-2024 · Mckesson Medical-Surgical Inc. Corporate Office

Hazard
Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result. · FDA Class II
Units
2 units
Sold at
US Nationwide distribution in the states of AZ, PA and TX.

Description

Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.

Product
Abbott Point of Care Control Control i-STAT¿ Level 3 1.7 mL
Manufacturer(s)
Mckesson Medical-Surgical Inc. Corporate Office

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov