Home / Recalls / Medical Devices
Medical Action Industries Inc recalls Max Barrier Kit, Sterile Single Use, Latex Free, 20/case, Medical Action Industries, Inc, Arden, NC.
Recalled July 9, 2014 · FDA recall Z-1954-2014 · Medical Action Industries
Hazard
Some of the sterile kits may have incomplete seals which might compromise the sterility of the kits. · FDA Class II
Units
474 cs (9,480 ea) total for all kits involved
Sold at
Distributed in the states of AR, AZ, FL, GA, IA, IL, KY, MA, MN, NC, NY, OH, TN, TX, and VA.
Description
Some of the sterile kits may have incomplete seals which might compromise the sterility of the kits.
Product
Max Barrier Kit, Sterile Single Use, Latex Free, 20/case, Medical Action Industries, Inc, Arden, NC.
Manufacturer(s)
Medical Action Industries Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →