Medical Components, Inc dba MedComp recalls 12F Tesio CATHETER SET (CUFF 32CM FROM TIP), REF MCTC1247SD, UDI 884908027266 Product Usage: The Medcomp Repair Kit

Recalled March 28, 2018 · FDA recall Z-1124-2018 · Medical Components, Inc

Hazard
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli. · FDA Class II
Units
N/A
Sold at
Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.

Description

There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.

Product
12F Tesio CATHETER SET (CUFF 32CM FROM TIP), REF MCTC1247SD, UDI 884908027266 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
Manufacturer(s)
Medical Components, Inc dba MedComp