Home / Recalls / Medical Devices
Medical Components, Inc dba MedComp recalls 12F X 60CM TESIO CATHETER SET WITH "Y" ADAPTOR, REF CMD757, UDI 884908005554 Product Usage: The Medcomp Repair Kit i
Recalled March 28, 2018 · FDA recall Z-1121-2018 · Medical Components, Inc
Hazard
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli. · FDA Class II
Units
55 kits
Sold at
Worldwide Distribution - US Nationwide in the states of AL, AR, CA, CO, CT, DE, FL, GA, IL, IN, KS, LA, MA, MD, MI, MN, MS, NC, NH, NJ, NM, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, VT, WA & WI. International: Australia, Austria, Belgium, Canada, Ecuador, France, Germany, Great Britain Greece, Ireland, Israel, Italy, Netherlands, Netherlands, South Africa, Spain Sweden, Turkey & United Arab Emirates.
Description
There is the potential for connection failures related to the female two-part luer to hemodialysis blood tubing lines, etc. which may contribute to blood loss or air emboli.
Product
12F X 60CM TESIO CATHETER SET WITH "Y" ADAPTOR, REF CMD757, UDI 884908005554 Product Usage: The Medcomp Repair Kit is indicated for use in replacing damaged female her connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.
Manufacturer(s)
Medical Components, Inc dba MedComp
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →