Home / Recalls / Medical Devices
Medical Components, Inc dba MedComp recalls Double Lumen CRRT/ABP Catheter Set; Hemodialysis catheter insertion kits, DL 11/15 DL 11/20 DL 9/15
Recalled September 5, 2018 · FDA recall Z-2933-2018 · Medical Components, Inc
Hazard
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death. · FDA Class II
Units
N/A
Sold at
Nationwide distribution; worldwide distribution.
Description
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Product
Double Lumen CRRT/ABP Catheter Set; Hemodialysis catheter insertion kits, DL 11/15 DL 11/20 DL 9/15
Manufacturer(s)
Medical Components, Inc dba MedComp
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →