RecallWatchUSA

Home / Recalls / Medical Devices

Medical Components, Inc dba MedComp recalls Duo-Flow XTP Straight Full Set; Hemodialysis catheter insertion kits, AAC02106 AAC02107 AAC02108

Recalled September 5, 2018 · FDA recall Z-2932-2018 · Medical Components, Inc

Hazard
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death. · FDA Class II
Units
N/A
Sold at
Nationwide distribution; worldwide distribution.

Description

The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.

Product
Duo-Flow XTP Straight Full Set; Hemodialysis catheter insertion kits, AAC02106 AAC02107 AAC02108
Manufacturer(s)
Medical Components, Inc dba MedComp