Medical Components, Inc dba MedComp recalls Medcomp Duo-Flow 400XL Catheter, Hemo-Flow Catheter, SLX Catheter, Triple Lumen Infusion Catheter, CT rated PICC, Split

Recalled April 8, 2015 · FDA recall Z-1351-2015 · Medical Components, Inc

Hazard
Medcomp has initiated the recall of Duo-Flow 400XL Catheter, Hemo-Flow Catheter, SLX Catheter, Triple Lumen Infusion Catheter, CT rated PICC, Split Cath III, PC Split Cath III because the product shipped was out of specification. One BI (biological indicator) out of 21 on the sterilization load was out of specification. · FDA Class II
Units
3152
Sold at
Worldwide Distribution.

Description

Medcomp has initiated the recall of Duo-Flow 400XL Catheter, Hemo-Flow Catheter, SLX Catheter, Triple Lumen Infusion Catheter, CT rated PICC, Split Cath III, PC Split Cath III because the product shipped was out of specification. One BI (biological indicator) out of 21 on the sterilization load was out of specification.

Product
Medcomp Duo-Flow 400XL Catheter, Hemo-Flow Catheter, SLX Catheter, Triple Lumen Infusion Catheter, CT rated PICC, Split Cath III, PC Split Cath III; AngioDynamics product- Dura-Flow Catheter, Schon XL Catheter; The affected product are insertion kits used to implant the devices packaged-long and short term hemodialysis catheters and infusion catheters.
Manufacturer(s)
Medical Components, Inc dba MedComp

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov