RecallWatchUSA

Home / Recalls / Medical Devices

Medicina Uk Ltd recalls Medicina ENFit 1ml Reusable Low Dose Enteral Syringe Code: LHE01LD

Recalled October 26, 2022 · FDA recall Z-0107-2023 · Medicina Uk

Hazard
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification · FDA Class II
Units
9 boxes (900 syringes)
Sold at
Nationwide

Description

Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification

Product
Medicina ENFit 1ml Reusable Low Dose Enteral Syringe Code: LHE01LD
Manufacturer(s)
Medicina Uk Ltd
Made in
United Kingdom

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Medicina Uk Ltd recalls Medicina ENFit 1ml Reusable Low Dose Enteral Syringe Code: LHE01LD | RecallWatchUSA