Home / Recalls / Medical Devices
Medicina Uk Ltd recalls Medicina Oral Tip Syringe 0.5ml Code: OT005
Recalled October 26, 2022 · FDA recall Z-0109-2023 · Medicina Uk
Hazard
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification · FDA Class II
Units
120 boxes (12,000 Syringes)
Sold at
Nationwide
Description
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification
Product
Medicina Oral Tip Syringe 0.5ml Code: OT005
Manufacturer(s)
Medicina Uk Ltd
Made in
United Kingdom
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →