RecallWatchUSA

Home / Recalls / Medical Devices

Medicina Uk Ltd recalls Medicina Oral Tip Syringe 0.5ml Code: OT005

Recalled October 26, 2022 · FDA recall Z-0109-2023 · Medicina Uk

Hazard
Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification · FDA Class II
Units
120 boxes (12,000 Syringes)
Sold at
Nationwide

Description

Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification

Product
Medicina Oral Tip Syringe 0.5ml Code: OT005
Manufacturer(s)
Medicina Uk Ltd
Made in
United Kingdom

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov