RecallWatchUSA

Home / Recalls / Medical Devices

Medline Industries Inc recalls Quick Release Gait Belt, 60

Recalled October 14, 2020 · FDA recall Z-0055-2021 · Medline Industries

Hazard
The buckle on the Gait Belt may break due to degradation of the material due to exposure to gamma radiation. · FDA Class II
Units
13,500 units
Sold at
Domestic Distribution Only: IL, MI

Description

The buckle on the Gait Belt may break due to degradation of the material due to exposure to gamma radiation.

Product
Quick Release Gait Belt, 60
Manufacturer(s)
Medline Industries Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov