Home / Recalls / Medical Devices
Medline Industries, LP recalls Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035; 2) KIT SURG ONC INSERT CENT.VENOU, Model Numb
Recalled February 25, 2026 · FDA recall Z-1402-2026 · Medline Industries
Hazard
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. · FDA Class II
Units
684 units
Sold at
Worldwide distribution - US Nationwide.
Description
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Product
Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035; 2) KIT SURG ONC INSERT CENT.VENOU, Model Number: DYKMBNDL43
Manufacturer(s)
Medline Industries, LP
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →