Home / Recalls / Medical Devices
Medline Industries, LP recalls Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ENDOSCOPY-LF, Medline Kit N
Recalled May 6, 2026 · FDA recall Z-2024-2026 · Medline Industries
Hazard
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin. · FDA Class II
Units
2153 kits
Sold at
Worldwide distribution.
Description
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Product
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ENDOSCOPY-LF, Medline Kit Number/SKU DYNJ22890M; 2. ENDOSCOPY PACK, Medline Kit Number/SKU DYNJ40645B; 3. ENT PK 1008847, Medline Kit Number/SKU DYNJ42319C; 4. ENDO SINUS PACK, Medline Kit Number/SKU DYNJ66449; 5. ENT, Medline Kit Number/SKU DYNJ900282I; 6. SEPTO/ENDO, Medline Kit Number/SKU DYNJ902831D; 7. ENDOSCOPIC SKULL BASE, Medline Kit Number/SKU DYNJ906630.
Manufacturer(s)
Medline Industries, LP
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →