RecallWatchUSA

Home / Recalls / Medical Devices

MEDLINE INDUSTRIES, LP - Northfield recalls Medline convenience kits containing semi-rigid suction liners labeled as follows: a) ENDO BRONCH PACK, Pack Number DYKE

Recalled September 18, 2024 · FDA recall Z-3111-2024 · MEDLINE INDUSTRIES, LP - Northfield

Hazard
Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use. · FDA Class II
Units
6257 units
Sold at
Worldwide distribution - US Nationwide and the countries of Canada, Panama.

Description

Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.

Product
Medline convenience kits containing semi-rigid suction liners labeled as follows: a) ENDO BRONCH PACK, Pack Number DYKE1355; b) ENDO BASIC KIT, Pack Number DYKE1356; c) ENDOSCOPY KIT, Pack Number DYKE1771C
Manufacturer(s)
MEDLINE INDUSTRIES, LP - Northfield