RecallWatchUSA

Home / Recalls / Medical Devices

MEDLINE INDUSTRIES, LP - Northfield recalls Medline procedure kits, labeled as: 1) WOUND CARE TRAY, REF DYNDA1412A; 2) LACERATION TRAY, REF DYNDL1263A.

Recalled May 7, 2025 · FDA recall Z-1711-2025 · MEDLINE INDUSTRIES, LP - Northfield

Hazard
It was identified that the Medline Sterile Water and Medline Sterile 0.9% Normal Saline bottles were leaking at the interface of the peel-foil and bottle opening. · FDA Class II
Units
5200 units
Sold at
Worldwide distribution - US Nationwide and the countries of Panama and Canada.

Description

It was identified that the Medline Sterile Water and Medline Sterile 0.9% Normal Saline bottles were leaking at the interface of the peel-foil and bottle opening.

Product
Medline procedure kits, labeled as: 1) WOUND CARE TRAY, REF DYNDA1412A; 2) LACERATION TRAY, REF DYNDL1263A.
Manufacturer(s)
MEDLINE INDUSTRIES, LP - Northfield

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

MEDLINE INDUSTRIES, LP - Northfield recalls Medline procedure kits, labeled as: 1) WOUND CARE TRAY, REF DYNDA1412A; 2) LACERATION TRAY, REF DYNDL1263A. | RecallWatchUSA