RecallWatchUSA

Home / Recalls / Medical Devices

MEDLINE INDUSTRIES, LP - Northfield recalls Medline Sterile Laparoscopy Pack, DYNJ68187A (REF), containing the component ConMed/Unimax Detachable Endo Pocket

Recalled October 16, 2024 · FDA recall Z-0053-2025 · MEDLINE INDUSTRIES, LP - Northfield

Hazard
CONMED Corporation (vendor) initiated a recall for select lot numbers of the Unimax Detachable Endo Pocket. The Detachable Endo Pocket is indicated for use as a receptacle for the collection and extraction of tissues, organs, and calculi during general and laparoscopic procedures. The tube inside the pouch, which detaches during the removal process, may stretch out from the pouch opening if it is not precisely fixed during the production process. Medline has identified that this product is utilized as a component within the Medline-branded convenience kits. · FDA Class II
Units
72 kits
Sold at
US Nationwide distribution in the state of CA.

Description

CONMED Corporation (vendor) initiated a recall for select lot numbers of the Unimax Detachable Endo Pocket. The Detachable Endo Pocket is indicated for use as a receptacle for the collection and extraction of tissues, organs, and calculi during general and laparoscopic procedures. The tube inside the pouch, which detaches during the removal process, may stretch out from the pouch opening if it is not precisely fixed during the production process. Medline has identified that this product is utilized as a component within the Medline-branded convenience kits.

Product
Medline Sterile Laparoscopy Pack, DYNJ68187A (REF), containing the component ConMed/Unimax Detachable Endo Pocket
Manufacturer(s)
MEDLINE INDUSTRIES, LP - Northfield

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

MEDLINE INDUSTRIES, LP - Northfield recalls Medline Sterile Laparoscopy Pack, DYNJ68187A (REF), containing the component ConMed/Unimax Detachable Endo Pocket | RecallWatchUSA