Medline Industries, LP recalls See RES for complete list. Medline Convenience Kits: 1) MAJOR LAPAROTOMY CDS, Model Number: CDS860015U; 2) MAJOR LAPAROT
Recalled May 13, 2026 · FDA recall Z-2098-2026 · Medline Industries
Hazard
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. · FDA Class II
Units
114778 units
Sold at
Worldwide - US Nationwide distribution.
Description
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Product
See RES for complete list. Medline Convenience Kits: 1) MAJOR LAPAROTOMY CDS, Model Number: CDS860015U; 2) MAJOR LAPAROTOMY CDS, Model Number: CDS860015V; 3) MAJOR LAPAROTOMY CDS, Model Number: CDS860015W; 4) MAJOR LAPAROTOMY CDS, Model Number: CDS860015X; 5) MINOR LAPAROTOMY CDS, Model Number: CDS860016P; 6) MINOR LAPAROTOMY CDS, Model Number: CDS860016Q; 7) MINOR LAPAROTOMY CDS, Model Number: CDS860016R; 8) MINOR LAPAROTOMY CDS, Model Number: CDS860016S; 9) LAP GASTRIC BYPASS CDS, Model Number: CDS860146M;
Manufacturer(s)
Medline Industries, LP