Home / Recalls / Medical Devices
Medline Industries, LP recalls See RES for complete list. Medline Convenience Kits: 1) PLASTIC BIN W/MISC KITS, Model Number: ACC010499; 2) BREAST-HERN
Recalled May 13, 2026 · FDA recall Z-2099-2026 · Medline Industries
Hazard
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. · FDA Class II
Units
5497 units
Sold at
Worldwide - US Nationwide distribution.
Description
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Product
See RES for complete list. Medline Convenience Kits: 1) PLASTIC BIN W/MISC KITS, Model Number: ACC010499; 2) BREAST-HERNIA-PORT CDS-LF, Model Number: CDS984853J; 3) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117C; 4) KIT PLASTIC INSERT TISSUE EXPA, Model Number: DYKMBNDL158; 5) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170; 6) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170A; 7) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170B; 8) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182;
Manufacturer(s)
Medline Industries, LP
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →