Medtronic Advanced Energy, LLC recalls Medtronic Advanced Energy Aquamantys3 Pump Generator; Product Catalog Number: 10-1357 (Generator assembly), 40-404-1(as
Recalled February 12, 2014 · FDA recall Z-0905-2014 · Medtronic Advanced Energy
Hazard
The internal protection circuitry on the electrosurgical output that prevents a patient from exposure to DC voltage when the device is activated is being compromised during normal use. · FDA Class II
Units
242 devices
Sold at
Nationwide Distribution including CA, MO, PA, NJ, NY, OH, VA, TX, NC, DE, KS, TN, LA, IL, MI, SC, WI, and FL.
Description
The internal protection circuitry on the electrosurgical output that prevents a patient from exposure to DC voltage when the device is activated is being compromised during normal use.
Product
Medtronic Advanced Energy Aquamantys3 Pump Generator; Product Catalog Number: 10-1357 (Generator assembly), 40-404-1(as shipped, including accessories) The Aquamantys3 System combines radio frequency (RF) energy and saline to reduce blood loss during and after surgical procedures. This patented Transcollation technology has been shown to reduce transfusion rates and may also reduce the need for other blood management procedures.
Manufacturer(s)
Medtronic Advanced Energy, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →