Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls CareLink SmartSync Device Manager - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P

Recalled May 12, 2021 · FDA recall Z-1528-2021 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Hazard
A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted. · FDA Class II
Units
16,311 units
Sold at
Worldwide distribution.

Description

A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.

Product
CareLink SmartSync Device Manager - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure, Model Number 24970A
Manufacturer(s)
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)