Home / Recalls / Medical Devices

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Cobalt XT HF CRT-D, M

Recalled August 17, 2022 · FDA recall Z-1486-2022 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Hazard
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds). · FDA Class I
Units
8429 units
Sold at
Worldwide Distribution

Description

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Product
Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Cobalt XT HF CRT-D, Model Numbers: DTPA2D4, DTPA2D1; b. Cobalt XT HF Quad CRT-D, Model Numbers: DTPA2QQ, DTPA2Q1
Manufacturer(s)
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov