RecallWatchUSA

Home / Recalls / Medical Devices

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Cobalt HF CRT-D, Mode

Recalled August 17, 2022 · FDA recall Z-1488-2022 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Hazard
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds). · FDA Class I
Units
1408 units
Sold at
Worldwide Distribution

Description

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Product
Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Cobalt HF CRT-D, Model Numbers: DTPB2D4, DTPB2D1; b. Cobalt HF Quad CRT-D, Model Numbers: DTPB2QQ, DTPB2Q1
Manufacturer(s)
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Cobalt HF CRT-D, Mode | RecallWatchUSA