Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Crome Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Crome HF CRT-D, Model
Recalled August 17, 2022 · FDA recall Z-1490-2022 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Hazard
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds). · FDA Class I
Units
866 units
Sold at
Worldwide Distribution
Description
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
Product
Crome Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Crome HF CRT-D, Model Numbers: DTPC2D4, DTPC2D1; b. Crome HF Quad CRT-D, Model Numbers: DTPC2QQ, DTPC2Q1
Manufacturer(s)
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →