Home / Recalls / Medical Devices
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls CRT-D DTBB1QQ VIVA QUAD S IS4/DF4 US, Model Number DTBB1QQ; Implantable Cardioverter Defibrillators
Recalled June 28, 2023 · FDA recall Z-1744-2023 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Hazard
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds. · FDA Class I
Units
193 units
Sold at
US Nationwide - Worldwide Distribution
Description
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Product
CRT-D DTBB1QQ VIVA QUAD S IS4/DF4 US, Model Number DTBB1QQ; Implantable Cardioverter Defibrillators
Manufacturer(s)
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →