RecallWatchUSA

Home / Recalls / Medical Devices

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Medtronic CareLink 2090 Programmer

Recalled November 14, 2018 · FDA recall Z-0406-2019 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Hazard
Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download. · FDA Class II
Units
30172 units
Sold at
Worldwide

Description

Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download.

Product
Medtronic CareLink 2090 Programmer
Manufacturer(s)
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Medtronic CareLink 2090 Programmer | RecallWatchUSA