RecallWatchUSA

Home / Recalls / Medical Devices

Medtronic Inc. Cardiac Rhythm Disease Management recalls Medtronic Affinity Pediatric Arterial Filter, model number S4014, sterilized using ethylene oxide. Recall Update 9/

Recalled August 21, 2013 · FDA recall Z-1966-2013 · Medtronic Inc. Cardiac Rhythm Disease Management

Hazard
Medtronic is recalling 148 Affinity Pediatric Arterial Filters from 2 manufacturing lots because a limited quantity of filters may have a small breach in the filter media that may allow unwanted particulate or gaseous emboli to pass through the filter and the outlet of the device, which could result in serious injury to a pediatric patient including neurological damage, or potential patient death. · FDA Class II
Units
153 devices.
Sold at
Worldwide Distribution - US Distribution only to MS and IL, and the countries of Argentina, South Africa and Singapore.

Description

Medtronic is recalling 148 Affinity Pediatric Arterial Filters from 2 manufacturing lots because a limited quantity of filters may have a small breach in the filter media that may allow unwanted particulate or gaseous emboli to pass through the filter and the outlet of the device, which could result in serious injury to a pediatric patient including neurological damage, or potential patient death.

Product
Medtronic Affinity Pediatric Arterial Filter, model number S4014, sterilized using ethylene oxide. Recall Update 9/6/13: Medtronic, Red cross Infant Pack 1/B, South-Africa, REF M272403C, Sterile EO. The Affinity Pediatric Arterial Filters are single-use, sterile, nonpyrogenic devices designed to filter microemboli greater than the specified micron size from the circuit for periods up to 6 hours during cardiopulmonary bypass surgery. These devices are available both in an uncoated and a Carmeda coated option. The Carmeda coating is a BioActive surface that is non-leaching and provides a thromboresistant blood contact surface.
Manufacturer(s)
Medtronic Inc. Cardiac Rhythm Disease Management

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Medtronic Inc. Cardiac Rhythm Disease Management recalls Medtronic Affinity Pediatric Arterial Filter, model number S4014, sterilized using ethylene oxide. Recall Update 9/ | RecallWatchUSA