RecallWatchUSA

Home / Recalls / Medical Devices

Medtronic Inc. recalls Medtronic Implantable Insulin Pump System, REF: MMT-2007D, Sterile EO, CE 0459 UDI: 00763000043001

Recalled June 10, 2020 · FDA recall Z-2206-2020 · Medtronic

Hazard
Internal testing results showed endotoxin-related test result did not meet specifications for implantable insulin pump. If pumps have higher than allowed endotoxin levels, patients may have an effect of fever, hypotension, anaphylactic shock, hyperglycemia, or diabetic ketoacidosis (DKA), potentially followed by secondary multi-organ failure (primarily renal and hepatic) and/or death. · FDA Class II
Units
24 Pumps
Sold at
No US Distribution. OUS: Netherlands, France, and Sweden

Description

Internal testing results showed endotoxin-related test result did not meet specifications for implantable insulin pump. If pumps have higher than allowed endotoxin levels, patients may have an effect of fever, hypotension, anaphylactic shock, hyperglycemia, or diabetic ketoacidosis (DKA), potentially followed by secondary multi-organ failure (primarily renal and hepatic) and/or death.

Product
Medtronic Implantable Insulin Pump System, REF: MMT-2007D, Sterile EO, CE 0459 UDI: 00763000043001
Manufacturer(s)
Medtronic Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov