RecallWatchUSA

Home / Recalls / Medical Devices

Medtronic Inc recalls Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MUS0130069X6

Recalled November 29, 2023 · FDA recall Z-0320-2024 · Medtronic

Hazard
The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side. · FDA Class II
Units
2767 units
Sold at
Worldwide distribution - US Nationwide and the countries of Algeria, Armenia, France, Germany, Iraq, Italy, Japan, Netherlands, Poland, Romania, Spain, Sweden, Switzerland, United Kingdom.

Description

The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side.

Product
Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MUS0130069X6
Manufacturer(s)
Medtronic Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Medtronic Inc recalls Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MUS0130069X6 | RecallWatchUSA