Medtronic Inc recalls Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MOM0130068X5
Recalled November 29, 2023 · FDA recall Z-0322-2024 · Medtronic
Hazard
The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side. · FDA Class II
Units
48 units
Sold at
Worldwide distribution - US Nationwide and the countries of Algeria, Armenia, France, Germany, Iraq, Italy, Japan, Netherlands, Poland, Romania, Spain, Sweden, Switzerland, United Kingdom.
Description
The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side.
Product
Medtronic Mo.Ma Ultra Proximal Cerebral Protection Device, Model Number MOM0130068X5
Manufacturer(s)
Medtronic Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →