Home / Recalls / Medical Devices
Medtronic Inc. recalls Medtronic REF: MMT-7020B, Guardian Sensor (3) for the following packaging: GS3 1PK 2L US/GTIN: 00763000140779 GS
Recalled May 26, 2021 · FDA recall Z-1602-2021 · Medtronic
Hazard
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings. · FDA Class II
Units
302,667 units
Sold at
Worldwide distribution.
Description
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
Product
Medtronic REF: MMT-7020B, Guardian Sensor (3) for the following packaging: GS3 1PK 2L US/GTIN: 00763000140779 GS3 1PK 2L US/ GTIN: 00763000337049 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.
Manufacturer(s)
Medtronic Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →