Medtronic MiniMed Inc. recalls CareLink iPro2 Reports, Model No. MMT-7340 is software intended to retrospectively calibrates sensor data and provides
Recalled April 27, 2016 · FDA recall Z-1501-2016 · Medtronic MiniMed
Hazard
Medtronic MiniMed is recalling 19 CareLink iPro Therapy Management reports because the CareLink iPro reports were created using the Pacific Standard Time zone as the time reference, although the patients and their sensor data were subject to the Central Standard Time zone. The CareLink iPro2 Professional CGM (continuous glucose monitor) data uploading error that resulted in an inaccurate time zone being applied to 19 CareLink iPro reports. · FDA Class II
Units
20
Sold at
US: MN
Description
Medtronic MiniMed is recalling 19 CareLink iPro Therapy Management reports because the CareLink iPro reports were created using the Pacific Standard Time zone as the time reference, although the patients and their sensor data were subject to the Central Standard Time zone. The CareLink iPro2 Professional CGM (continuous glucose monitor) data uploading error that resulted in an inaccurate time zone being applied to 19 CareLink iPro reports.
Product
CareLink iPro2 Reports, Model No. MMT-7340 is software intended to retrospectively calibrates sensor data and provides reports of continuous glucose information
Manufacturer(s)
Medtronic MiniMed Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →