Home / Recalls / Medical Devices
Medtronic Navigation, Inc. recalls Medtronic Nexframe Stereotactic System and StealthStation Cranial software version 3.0 or newer with StealthStation DBS
Recalled May 27, 2020 · FDA recall Z-1995-2020 · Medtronic Navigation
Description
Entry point and lead placement inaccuracies during deep brain stimulation lead implantation procedures may occur when using a specific combination of the firm's Steriotactic System and auto-registration feature with a specific imaging system (also known as a fiducial-less procedure). Minor patient movement may not be initially detected by the user or the software during the auto-registration scan process potentially resulting in inaccuracies and risks for the patient including: inaccurate lead placement, delay of surgery, aborted surgery, or additional intervention (including revision of the lead placement and subsequent imaging).
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →