Medtronic Navigation, Inc. recalls StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systems
Recalled June 14, 2023 · FDA recall Z-1666-2023 · Medtronic Navigation
Hazard
In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury · FDA Class I
Units
556
Sold at
US Distribution: PR. OUS (foreign) distribution: ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BOSNIA and HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CANARY ISLANDS, CHILE, Colombia COLOMBIA, C¿TE D'IVOIRE, CROATIA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, HUNGARY, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KENYA, LATVIA, MEXICO, MOLDAVIA, NETHERLANDS, PAKISTAN, POLAND, PORTUGAL, REPUBLIC of KOREA, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UNITED ARAB EMIRATES
Description
In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury
Product
StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systems
Manufacturer(s)
Medtronic Navigation, Inc.