RecallWatchUSA

Home / Recalls / Medical Devices

Medtronic Navigation, Inc. recalls StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation System, with StealthStation Spin

Recalled October 16, 2024 · FDA recall Z-0050-2025 · Medtronic Navigation

Hazard
Due to two software anomalies that can occur within a spine procedure. Two issues are: 1. Navigation Orientation - when lateral orientation radiographic images are taken, information may appear flipped when utilizing the system (180 degree flipped). 2. Pop-up Message - when using CT+Fluouroscopy images with a slice spacing thickness greater than 2mm, the system does not display a pop-up message indicating the slice spacing is not optimal. · FDA Class II
Units
1,801 systems
Sold at
Worldwide - U.S Nationwide distribution including in the states of AK, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, TN, TX, UT, VA, WA, WI and WV. The countries of Argentina, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, Brazil, BULGARIA, CANADA, CHILE, China, Colombia, Costa Rica, CROATIA, CYPRUS, DENMARK, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, HUNGARY, ICELAND, India, ISRAEL, ITALY, Japan, Jordan, KAZAKHSTAN, KENYA, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, Malaysia, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, OMAN, PAKISTAN, PANAMA, Philippines, POLAND, PUERTO RICO, QATAR, Republic of Korea, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, Singapore, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, Turkey, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM.

Description

Due to two software anomalies that can occur within a spine procedure. Two issues are: 1. Navigation Orientation - when lateral orientation radiographic images are taken, information may appear flipped when utilizing the system (180 degree flipped). 2. Pop-up Message - when using CT+Fluouroscopy images with a slice spacing thickness greater than 2mm, the system does not display a pop-up message indicating the slice spacing is not optimal.

Product
StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation System, with StealthStation Spine software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical and orthopedic procedures.
Manufacturer(s)
Medtronic Navigation, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Medtronic Navigation, Inc. recalls StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation System, with StealthStation Spin | RecallWatchUSA