RecallWatchUSA

Home / Recalls / Medical Devices

Medtronic Perfusion Systems recalls Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1

Recalled March 4, 2026 · FDA recall Z-1473-2026 · Medtronic Perfusion Systems

Hazard
During the manufacturing, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design. · FDA Class II
Units
939 units
Sold at
US Nationwide. Global Distribution.

Description

During the manufacturing, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.

Product
Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1
Manufacturer(s)
Medtronic Perfusion Systems

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Medtronic Perfusion Systems recalls Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product Number TSMICS1 | RecallWatchUSA