RecallWatchUSA

Home / Recalls / Medical Devices

Medtronic Sofamor Danek USA Inc recalls Medtronic ZEVO(TM) Anterior Cervical Plate System SLOT SCREW, REF 7710015, Size 4.5mm x 15mm Product Appliance, fixat

Recalled May 16, 2018 · FDA recall Z-1750-2018 · Medtronic Sofamor Danek USA

Hazard
Slot screw had an overall length of 15mm, but per the specification, the screw should have an overall length of approximately 17mm. · FDA Class II
Units
39 units
Sold at
Worldwide Distribution - US Nationwide in the states of GA, CT, LA, MD, WY, IN, MD, FL, and France

Description

Slot screw had an overall length of 15mm, but per the specification, the screw should have an overall length of approximately 17mm.

Product
Medtronic ZEVO(TM) Anterior Cervical Plate System SLOT SCREW, REF 7710015, Size 4.5mm x 15mm Product Appliance, fixation, spinal intervertebral body
Manufacturer(s)
Medtronic Sofamor Danek USA Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov