Home / Recalls / Medical Devices
MedX Holdings, Inc. recalls Cervical Extension Machine for Physical Therapy
Recalled January 14, 2015 · FDA recall Z-0936-2015 · MedX Holdings
Hazard
Review of DHRs revealed that HiPot, Continuity, load and friction re-testing was not completed on these devices during manufacturing. · FDA Class II
Units
4
Sold at
Worldwide Distribution - USA including MO, IL, ID, and MN and Internationally to Germany and South Korea.
Description
Review of DHRs revealed that HiPot, Continuity, load and friction re-testing was not completed on these devices during manufacturing.
Product
Cervical Extension Machine for Physical Therapy
Manufacturer(s)
MedX Holdings, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →