Merck Millipore Ltd. recalls Minicon Clinical Sample Concentrators for In Vitro Diagnostic Use 9031 (Model: Minicon B-15) & 9051 (Model: Minicon CS-1

Recalled November 28, 2018 · FDA recall Z-0489-2019 · Merck Millipore

Hazard
In vitro diagnostic devices, used greater than 6 months after date of manufacture, may have increased concentration times of greater than the specification of 3 hours and up to 24 hours, which may lead to delays in carrying out the subsequent analysis tests. · FDA Class II
Units
1360 units
Sold at
Worldwide Distribution: US (nationwide) to states of: AZ, CA, CT, DC, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NJ, NY, OH, OK,OR, PA, RI, TN, TX, VA, VT, WA and WI; and countries of: Canada and Australia.

Description

In vitro diagnostic devices, used greater than 6 months after date of manufacture, may have increased concentration times of greater than the specification of 3 hours and up to 24 hours, which may lead to delays in carrying out the subsequent analysis tests.

Product
Minicon Clinical Sample Concentrators for In Vitro Diagnostic Use 9031 (Model: Minicon B-15) & 9051 (Model: Minicon CS-15) Minicon concentrators are non-sterile, disposable, multiwall ultrafiltration devices. Minicon¿ clinical sample concentrators are for in vitro diagnostic use and are intended for concentrating serum, urine, cerebrospinal fluid, and other body fluids prior to analysis.
Manufacturer(s)
Merck Millipore Ltd.
Made in
Ireland