RecallWatchUSA

Home / Recalls / Medical Devices

Meridian Bioscience Inc recalls illumipro-10 Incubator/Reader, Meridian Bioscience, Inc. The illumipro-10 incubator/reader is designed to be used wit

Recalled July 9, 2014 · FDA recall Z-1941-2014 · Meridian Bioscience

Hazard
Illumipro-10 instruments may have Block B chambers operating at an incorrect temperature as a result of incorrect temperature calibration. · FDA Class II
Units
35 instruments
Sold at
Worldwide Distribution - USA including AL, AZ, CA, CO, FL, IA, IN, LA, MA, ME, MI, MN, NJ, NY, OH, PA, and WV. Internationally to Australia.

Description

Illumipro-10 instruments may have Block B chambers operating at an incorrect temperature as a result of incorrect temperature calibration.

Product
illumipro-10 Incubator/Reader, Meridian Bioscience, Inc. The illumipro-10 incubator/reader is designed to be used with illumigene Molecular Assays. The device incubates illumigene Test Devices and reads endpoint reactions.
Manufacturer(s)
Meridian Bioscience Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov