Home / Recalls / Medical Devices
Merit Medical Systems, Inc. recalls High Pressure Tubing: REF: 502211002, HP9720E/C, HP9721E/C
Recalled August 20, 2025 · FDA recall Z-2293-2025 · Merit Medical Systems
Hazard
High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient. · FDA Class II
Units
8040
Sold at
US Nationwide distribution in the states of OH, CA, SD, VA, AZ, MT, TX, MI, FL, TN, PA, NJ, MN, ND, IA, LA, UT, KY, IL, OK, ME, IN, NY, NM, SC, AL, ID, CT, WI, GA, MA, NC, AR, WA.
Description
High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.
Product
High Pressure Tubing: REF: 502211002, HP9720E/C, HP9721E/C
Manufacturer(s)
Merit Medical Systems, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →