Michigan Instruments, Inc. recalls Thumper Cardiopulmonary Resuscitator, Model 1007CCV, Catalog No. 15370 Product Usage: The device is used to perform Ca

Recalled May 16, 2018 · FDA recall Z-1632-2018 · Michigan Instruments

Hazard
A performance issue with the chest depth needle valve (CDNV) may arise where turning the knob will not deliver any compression depth to the piston during the first 5 revolutions of the knob.The last ¿ to ¿ revolution of the knob could deliver a full 8cm of compression. · FDA Class III
Units
38 devices
Sold at
Distributed to China.

Description

A performance issue with the chest depth needle valve (CDNV) may arise where turning the knob will not deliver any compression depth to the piston during the first 5 revolutions of the knob.The last ¿ to ¿ revolution of the knob could deliver a full 8cm of compression.

Product
Thumper Cardiopulmonary Resuscitator, Model 1007CCV, Catalog No. 15370 Product Usage: The device is used to perform Cardiopulmonary Resuscitation (CRP) on adult patients and only adult patients in cases of clinical death as defined by a lack of spontaneous breathing and pulse.
Manufacturer(s)
Michigan Instruments, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov