Home / Recalls / Medical Devices
Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71
Recalled December 18, 2024 · FDA recall Z-0665-2025 · Micro-X
Hazard
Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator. · FDA Class II
Units
18
Sold at
US Nationwide Distribution: NJ, CA, PR
Description
Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
Product
MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71
Manufacturer(s)
Micro-X Ltd.
Made in
Australia
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →