Microbiologics Inc recalls Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184.
Recalled February 21, 2024 · FDA recall Z-1075-2024 · Microbiologics
Hazard
The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the negative control. · FDA Class II
Units
580 packages
Sold at
US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NE, NH, NJ, NY, OH, PA, RI, TN, TX, VA, WA, WI, WV, and WY. There was government distribution and no military distribution. Foreign distribution was made to France.
Description
The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the negative control.
Product
Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184.
Manufacturer(s)
Microbiologics Inc