Microgenics Corp recalls DRI Cocaine Metabolite Assay, labeled as Multigent Cocaine Assay for Abbott Laboratories. 100 ml, IVD Mfg by: Micro

Recalled October 2, 2013 · FDA recall Z-2267-2013 · Microgenics

Hazard
Surveillance of reagent stability for specific lots of DRI Cocaine Metabolite Reagent show an unexpected decreasing change in the reaction rate, indicating a faster than expected reagent degradation. · FDA Class II
Units
1473 units in distribution.
Sold at
US Distribution including the states of DC, MA, NJ, TX. FL, CA, WA, NY, TX, CT, KY, LA, AR, TN, AL, GA, CT, VA, NC, KS, MD, OH and IL

Description

Surveillance of reagent stability for specific lots of DRI Cocaine Metabolite Reagent show an unexpected decreasing change in the reaction rate, indicating a faster than expected reagent degradation.

Product
DRI Cocaine Metabolite Assay, labeled as Multigent Cocaine Assay for Abbott Laboratories. 100 ml, IVD Mfg by: Microgenics Corporation, 46360 Fremont Blvd. Fremont, CA 94538 The DRI Cocaine Metabolite Enzyme Immunoassay is a homogeneous enzyme intended for the qualitative and semi-quantitative determination of benzoylecgonine (Cocaine Metabolite) in human urine with either 300 ng/mL or 150 ng/mL as a cutoff calibrator. The assay provided only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirm analytical result.
Manufacturer(s)
Microgenics Corp