Home / Recalls / Medical Devices
Microgenics Corporation recalls DRI Salicylate Serum Tox Assay, Catalog number 0977. Toxicology: The DRI Salicylate Serum Tox Assay is intended for
Recalled October 5, 2016 · FDA recall Z-2884-2016 · Microgenics
Hazard
Some lots of DRI Salicylate Serum Tox Assay negative patient samples are recovering higher than the product sensitivity claim of 4.4 mg/dL. · FDA Class II
Units
1708 kits
Sold at
Worldwide Distribution: US (nationwide) to states of: AZ, CA, MO, TX, and PR and countries of: Canada, Germany, Hong Kong, Italy, Malaysia, and Taiwan.
Description
Some lots of DRI Salicylate Serum Tox Assay negative patient samples are recovering higher than the product sensitivity claim of 4.4 mg/dL.
Product
DRI Salicylate Serum Tox Assay, Catalog number 0977. Toxicology: The DRI Salicylate Serum Tox Assay is intended for the quantitative determination of salicylate in human serum or plasma.
Manufacturer(s)
Microgenics Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →